Details of Calibration Field Probes
Calibration procedure
- IEEE Std 1309-2013
Our Facilities
- TEM-cells 5 Hz - 500 MHz, up to 300 V/m
- Anechoic chamber 100 MHz - 40 GHz, up to 200 V/m
- Reverberation Chamber (1 GHz – 4 GHz: 1.2 kV/m; 4 GHz – 12 GHz: 600V/m; 12 GHz – 18 GHz: 300V/m)
- Parallel-plate capacitor 0 - 10 kHz, up to 1 kV/m
- Magnetic field generators (coils) 0 - 10MHz, up to 400 A/m
Features
- Determination of the calibration factor as function of frequency, probe orientation, field strength and modulation
- Isotropy measurements as function of frequency
- Evaluation of amplitude linearity as function of frequency
The uncertainty (95%, k=2) of the established field is:
- ±5%, DC - 10 kHz
- ±15%, 10 kHz - 150 MHz
- ±20%, 150 MHz - 18 GHz
- ±25%, 18 GHz - 40 GHz