Publications Miscellaneous

Radio Frequency Engineering

2021 - SUPPLEMENTARY COMPARISON EURAMET supplementary comparison on calibration of RF current monitoring probe

O Şen, M Çınar, A Kriz, T Pavlíček, J M Lerat, M Wojciechowski, B Pinter, M Rodríguez and F Pythoud

Metrologia, Volume 58, Number 1A, published 28 June 2021, © 2021 BIPM & IOP Publishing Ltd

Abstract:
Introduction - RF current monitoring probes are commonly utilized in measurements of disturbance currents of cables without making direct conductive contact with the source conductor and without modification of its circuit. They must be calibrated by National Metrology Institutes (NMIs) or accredited calibration laboratories in accordance with international standards such as CISPR 16-1-2 or ISO 11452-4. The RF current monitor probe comparison measurements were conducted with 8 participants; five national metrology institutes (NMIs), two designated institutes (DI) and one commercial calibration laboratory at the frequencies 0.01 MHz, 0.1 MHz, 1 MHz, 10 MHz, 100 MHz, 200 MHz, 300 MHz, 400 MHz and 500 MHz in order to obtain the transfer impedance of the RF current monitoring probe. The comparison measurements started in September 2019 and were completed in June 2020.

 

>> Download Paper in Adobe Acrobat Format​​​​​​​

<< back to Publications

 


Download the Free Acrobat Reader